Efficiency Test Of Colimator Shutter At The X Ray Tube In Radiodiagnostic Laboratory Of Poltekkes Jakarta 2 And Two Clinical Hospitals In Jakarta
https://doi.org/10.36525/sanitas.2017.3
Keywords
shutter collimation Permenkes No. 1250/Menkes/SK/XII/2009 equipment x-rayAbstract
The purpose of this study was to determine the efficiency of the collimator shutter with the presence or absence of leakage from the collector slit of the x-ray equipment. The type of research used is qualitative method using descriptive cross sectional method, conducting collimatory slot testing against x-ray equipment in laboratory and two installation of Radiology field of clinical practice for student of Radiodiagnostic and Radiotherapy of PoltekkesKemenkes Jakarta II. Result, there was a leakage of collimation shutter with marked blackening on x-rayfilm, on 8 (eight) x-ray equipment, there was a full leakage (very significant), and there was a big leak, and there was a slight leakage. Conclusion, in this research there is one x-ray equipment according to the standard reference (golden standard), set by Permenkes No. 1250 / Menkes / SK / XII / 2009 that is SH x-ray equipment in Y hospital.
References
Carlton R. & AM. Principle of Radiographic Imaging. New York: Delmar Publicer; 2002.
Soekardjo M. Kajian Uji Kesesuaian Pesawat Sinar- X di Instalasi Radiologi. 2015;79–82.
Curry TS. Dowdey JE. & Murry RC. Christensen’s Physics of Diagnostik Radiology. 4th ed. Philadelphia: Lippincott Williams & Wilkins; 1990.
Fauber TL. Radiographic imaging & Exposure. 3rd ed. St Louis: Mosby, 2009
NRCP Report No.99. Quality Assurance For Diagnostic Imaging. Bethesda; 1995.
Peraturan pemerintah no. 63. Keselamatan dan Kesehatan terhadap radiasi pengion. Jakarta; 2000.
International Commision Radiation Protection(IRCP) No.103. Nilai Batas Dosis (NBD). 2007.
Akhadi Mukhlis. Dasar-dasar Proteksi Radiasi. Jakarta: PT Rineka Cipta; 2000.
BATAN. Peraturan Pemerintah Republik Indonesia Nomor 29 Tahun 2008 tentang Perizinan Pemanfaatan Sumber Radiasi Pengion dan Bahan Nuklir. 2008;(2).
BAPETEN. Uji Kesesuaian Pesawat Sinar-X Radiologi Diagnostik dan Intervensional. 2011;
Chesney DN & MO Radiographic Imaging. London: CBS Publicer; 2005.
Keputusan Menteri Kesehatan Republik Indonesia Nomor 1250/MENKES/SK/XII/. Pedoman Kendali Mutu (Quality Control) Peralatan Radiodiagnostik. Jakarta; 2009.